A single-channel biopotential recorder for educational demonstration. The analog front end uses discrete components for amplification and initial filtering. The system uses an ESP32-S3 MCU and an external Tang Nano 9K (GW1NR-9). The FPGA acquires ADC data, performs digital filtering (0.67 Hz HPF, selectable 40/100/200 Hz LPF, optional 50 Hz notch), and implements Pan–Tompkins R-wave detection. A DAC outputs waveforms for debugging. The MCU handles control, UI, and data export (CSV/BDF).

Version

1.0

Licenses
Hardware
CERN-OHL-P-2.0
Software
MIT
Documentation
CC-BY-4.0
View Documentation

Certification mark

The OSHW Certification Mark shows the project's unique certification ID and makes it easy for users to quickly find this page along with the project's details. This mark can be used throughout the project's design, documentation, and promotional material.

For more details, take a look at the Certification Mark Usage Guide.


Project Type
Science
Education
Electronics
Keywords
esp32 fpga analog biosignal eeg ecg emg bioamplifier pan-tompkins